- Glass autoclaves pressures upto 6 bar.
- 100 ml to 2 ltrs. volume.
- Ideal for corrosive applications and to view the reaction.
- High torque zero leakage magnetic drive coupling.
- Fully automated PC controlled high pressure system / pilot plant to continuously monitor, control & record various parameters like temperature, pressure, motor speed, gas / liquid flow, level, pH, turbidity, torque etc.
- Polycarbonate transparent cover is provided from all 4 sides of glass vessel for safety.
- Material of construction (MOC): Glass with SS316 / Peek / PFA coating / Hastelloy B/C / Monel / Inconel / Nickel / Titanium / Tantalum / Zirconium etc.
DESCRIPTION | STANDARD | OPTIONAL |
---|---|---|
Volume |
100 ml, 250 ml, 450 ml, 750 ml, 1 ltr., 2 ltrs. net filling capacity |
N.A. |
Material of construction (MOC) |
Borosilicate glass cylinder with top-bottom flange, head & other wetted parts of SS-316 |
Jacketed / non-jacketed glass with wetted parts of PTFE / PTFE coated SS316, Hastelloy C, Titanium, Monel, Inconel etc
|
Design Pressure |
6 bar (87 psi) |
10 bar (145 psi) upto 1 ltr. glass autoclave |
Design Temperature |
150°C |
200°C for jacketed vessels (by circulation of hot oil through jacket) |
Nozzles, Valves & Fittings |
Pressure gauge, vent needle valve, safety rupture disc, internal cooling (single tube in tube out type cooling pipe for 50 ml / 100 ml, serpentine cooling coil for 250 ml & helical cooling coil for 450 ml to 2 ltrs.), thermowell (with RTD-PT-100 temp. sensor), gas inlet & liquid sampling needle valve with common dip tube (sparger) |
Ball valve for powder / liquid inlet , serpentine cooling coil (450 ml -2 ltrs.) |
Body & Head Sealing |
Split clamp type quick opening system with clamp bolts & flat gasket |
Split taper clamp with bolts
|
Gasket |
PTFE |
N.A. |
Head Mounting Style |
Removable head and vessel design |
Fixed head design with removable vessel & heater (manual)- for 100 ml-1 ltr. size
|
Motor & Drive |
Non flame proof / Flameproof (Explosion proof) AC motor for group IIA & IIB gases |
Flame proof CE-Atex certified motor for group IIC gases like H2
|
Stirrer RPM |
100 – 1450 rpm variable speed |
2900 rpm |
Shaft Sealing |
Zero leakage Magnetic Drive coupling |
N.A. |
Stirre |
4 bladed pitch blade turbine (100 ml – 250 ml)
|
Hollow shaft with gas induction Impeller, propeller, anchor, spiral, ruston turbine, paddle & specially designed stirrer for highly viscous material |
Heating System |
Electric plate heater from bottom |
By circulation of hot oil through glass jacket |
Control Panel |
Compact SS – Control Panel with microprocessor based programmable PID temperature controller cum indicator, variable motor speed drive & other digital indicators/ controllers (as ordered) mounted on it |
Digital temperature indicator in flameproof enclosure with SS control panel mounted at a safe distance of 10m
|
Power Supply |
1Ø, 230 V AC, 50 Hz / 60 Hz |
1Ø, 110 V AC, 60 Hz
|
Mounting |
Table top (100 ml – 250 ml)
|
SS trolley (50 ml – 250 ml)
|
Safety Shield |
Polycarbonate transparent cover provided for safety in case of breakage |
N.A. |
Technical Specifications for Glass with jacket system
SR.NO. |
AUTOCLAVEVOLUME |
VESSEL I/D(MM) |
VESSEL INTERNALHEIGHT (MM) |
AC MOTOR(HP) |
OVERALL DIMENSIONS TABLE TOP MODEL WITHOUT PANELW X D X H (MM) |
OVERALL DIMENSIONSTROLLEY MODELW X D X H (MM) |
POWDER/LIQUIDINLET SIZE (INCH) |
FLUSH BOTTOMVALVE DIA (MM) |
---|---|---|---|---|---|---|---|---|
1 |
100 ml |
45 |
90 |
1/4 |
275 x 230 x 750 |
1010 x 410 x 1500 |
N.A. |
N.A. |
2 |
250 ml |
63 |
90 |
1/4 |
275 x 230 x 750 |
1010 x 410 x 1500 |
N.A. |
N.A. |
3 |
450 ml |
77 |
110 |
1/4 |
320 x 270 x 900 |
1010 x 410 x 1650 |
1/4” |
10 |
4 |
750 ml |
77 |
185 |
1/4 |
320 x 270 x 975 |
1010 x 410 x 1725 |
1/4” |
10 |
5 |
1 ltr. |
104 |
165 |
1/4 |
365 x 310 x 925 |
1010 x 410 x 1675 |
1/4” |
10 |
6 |
2 ltrs. |
104 |
305 |
1/4 |
355 x 310 x 1050 |
1040 x 410 x 1800 |
1/4” |
10 |
Note:
- The above specifications are for design pressure of 6 bar & 150 °C temperature. The same may change for higher pressure / temperature.
- Overall dimensions are indicative, they may change depending on the optional accessories or specific design modifications opted.
- Maximum reaction volume will be 75% of net filling volume (i.e. autoclave volume).
- Total equipment electrical load = Motor kw + Auto Cooling pump (0.37kw) (if opted)+(any accessories if selected)
Since development is continuous process, the above specifications are subject to change without prior notice.